Communication between winIDEA and the BlueBox fails and generates an error. Once you execute the Communication test, it displays the message: Hardware not found. Possible causes and solutions BlueBox is turned off Turn on the BlueBox and try again. Wrong hardware type selected Verify the H...
These tips help you resolve issues when connecting winIDEA to a BlueBox via USB on Linux. 1. Verify environment variable WINEDLLPATH The folder containing winIDEA.exe must be included in the environment variable WINEDLLPATH . This is normally set automatically in the winidea.sh script whic...
Note that e very BlueBox must have a unique IP Address. If more BlueBoxes are connected to the Ethernet and have the same IP Address set, only one will be active. BlueBoxes using network connections must be located in the same Broadcast Domain. Possible causes and solutions IP Address is i...
Possible issues and solutions DAP/DAPE connector misconfiguration In case the target has a DAP/DAPE connector without a notch (keying shroud) , make sure you properly orient the connector. Pin 1 from the Active Probe must align with the pin 1 from the target. Make sure that signals on the ...
If options Open-drain RESET drive mode and Latch target RESET are enabled in Hardware | CPU Options | Reset , BlueBox will hold the RESET line low whenever an external RESET pulse is detected. On some targets, the RESET line may become noisy enough that the BlueBox falsely detects external...
Delivering deeply tested products is our quality objective, nevertheless, all BlueBox systems come with a warranty. TASKING provides a free exchange system or a free repair during the warranty period if the hardware malfunction is not the result of an overvoltage, mechanical damage, or dam...
The following cross-over tables list all available Converters and their target connection. The tables explain which BlueBox Debug Adapter or Active Probe the particular Converter can be used with. iC7 BlueBox Hardware Debuggers iC5 BlueBox Hardware Debuggers More resources inHardware User ...
When configuring a BlueBox device to obtain an IP address via DHCP, the IP address may not be assigned even though: Option Obtain IP address automatically (DHCP) is selected in Hardware | Hardware Type | System Configuration . The DHCP server exists on the same VLAN as the BlueBox, or a fi...
Company network policies can prevent connecting the debugger to the network via Ethernet cable. In such cases, f ollow the procedure below to directly connect the debugger to the PC's Ethernet port via Ethernet cable. Solution 1.Disconnect the PC from the local network and connect the debu...
Embedded development and test means working with a live target system. Connecting electronic hardware to development equipment, a PC, and other resources under development or test, poses some risk. This Technical Note describes recommended measures to reduce the risk of electrical injury a...
If you are experiencing issues with establishing debug session, follow the below tips: Possible solutions Power-cycle your device and try to perform a reset This way you will pinpoint the issue either with: Establishing debug session (if reset fails as well) or Flash programming (if reset ...
Emulation Adapters are configured by default to use the target board clock source. This configuration is set on jumpers J2 and J3 , or DIP switch DIP2 . The default position is set closer to the Pin 1 marking. However, when using a crystal circuit oscillator on the target board, it might l...
Possible solutions Confirm the correct BlueBox selection Ensure you are connected to the correct iC7max/iC5700 to which you have connected the Active Probe. Check via Hardware | Debugger Hardware . Confirm indicator light status The indicator LED on the Active Probe should light up or star...
If the embedded target does not provide a matching debug connector (e.g. has a different pinout as the BlueBox debug adapter), it is not recommended to use wires or any other material ( example 1 , example 2 , example 3 - note that these are examples of not working systems). Possible solut...
The orange LED (marked F) : It is expected to turn on when the BlueBox is waiting for communication with the PC and to turn off once the communication is established. It will be blinking during a firmware upgrade with the red LED marking the progression through the upgrade steps. Other tha...
The iC7 BlueBox Comparison table display s the features and differences of the iC7mini, iC7pro, and iC7max BlueBox. The iC5BlueBox Comparison table compares the iC5700 and iC5000 BlueBox.
Cypress Traveo II packages (64-pin, 80-pin, 100-pin, and 144-pin) have different trace capabilities. The following Emulation Adapters provide the trace interfaces, where a trace port is not available or lacks the full trace capability: CYT2B9 CYT2B7 CYT2BL By default, port pins P18 (P18_3 ...
If the DHCP server is unavailable or the BlueBox cannot be discovered because it is not within the same broadcast domain, you must manually assign the IP address. If the BlueBox is configured to acquire the IP address from a DHCP server, but this fails, it will use the IP configuration spe...
Solution After you have successfully connected hardware and selected the current range according to the shunt resistor (JB1 jumper block), and voltage range depending on your target power supply range ( JB2 jumper block ) on the Power Probe, configure the following settings in winIDEA . Mo...
You get a BlueBox, hook it up, and can’t start debugging? What went wrong and what can you do about it?This topic lists the most common reasons for failure that you as a first-time user can encounter. Possible causes and solutions Licensing issues If you have more iC5 BlueBox Debuggers, yo...
The Active Probe is connected to the target on one side and the other side it connects to the iC7max/iC5700 FNet port via the FNet cable. In the Setup and Configure Hardware Tutorial you'll find detailed steps to connect to the following Active Probes: Infineon DAP/DAPE II AGBT/SGBT/Aurora...
In certain network situations, BlueBox will not automatically receive an IP address. In such cases, it is possible to set a fixed IP address manually. To set up the TCP/IP-Connection manually the BlueBox will need to be connected temporarily to the development PC via a USB cable. Possible ...
The Debug Adapter is directly connected to the Target and the other side connects via the ribbon cable to the iC5 (iC5000/iC5700) or iC7 (iC7mini/iC7pro) connectors. Detailed steps in Hardware Setup and Configuration Tutorial. Configuration steps 1. Connect the Power supply. 2. Connect the...
For powering use only original accessories, which are delivered together with the BlueBox hardware. If you wish to use a power supply different from the delivered one, consult with technical support first. Powering On/Off steps 1. Power on the BlueBox. 2. Power on the Target Board. When de...
When trying to perform a firmware update, the update fails. The issue can manifest itself in a variety of ways, from winIDEA becoming unresponsive for a while to an error message stating Emulator Boot error , however the end result is always the same: when checking the firmware version, th...
Would you like to utilize a solid-state relay in combination with BlueBox and its accessories to gain control of the connected target's power source? Find out more in winIDEA Help - Target power control using BlueBox & ADIO .
When assembling the Solder part and Extender part make sure to: 1. Align parts with pin 1. All pins on the solder part must be aligned with the pins on the Extender. 2. Gently push down in the middle of the Extender in the direction of the Solder part. Why Extenders? Extenders can be stack...